20.2 Charging Station
1. Disconnect the charging station from the electric supply system.
2. Proceed as described in chapter 20.1, steps 3 to 5.
20.3 Test-Sensor
Cleaning and disinfection as described in chapter 20.1, steps 3 to 5.
20.4 Reusability
Frequent reprocessing has no effect or limitation on the system components of the Occlu-
Sense
®
-System, as the end of the product's life is determined by wear and tear and da-
mage caused by use.
The use of damaged and dirty components is the responsibility of the user.
In case of disregard any liability is excluded.
NOTE
The instructions of the manufacturer of the detergent and disinfectant must
be followed strictly. In particular, the concentrations to be used and the ap-
plication times must be observed!
CAUTION
The components of the OccluSense
®
-System are not suitable for mechanical
cleaning and thermal disinfection processes.
The battery compartment cover must be closed during cleaning and disinfection.
Switch off the handheld before cleaning/disinfection!
It is mandatory that the contact time of the disinfectant product is being obser-
ved and that the unit is NOT switched on during this time.
The OccluSense
®
-Sensors are disposable medical devices and must not be reu-
sed.