RECORDINGS
14 Start a recording
14.1 General
Before starting the recording, the following preparations have to be made:
- Create / select patient
- Insert the sensor into the handheld
- Select capture mode
An upright sitting position of the patient is recommended.
NOTE
If this has not yet been performed, the initial setup has to be made first (see •
chapter 10).
The recording will be started and stopped by the control buttons of the •
OccluSense
®
handheld only.
For recording, a permanent, wireless connection between the handheld and •
OccluSense
®
-iPad-App is required.
CAUTION
The handheld may only be used with closed access covers (battery compartment •
and sensor lid) AND inserted sensor in the patient environment.
Replacing the sensors or batteries which require to open the respective access •
covers must be done outside the patient environment.
The practitioner must not touch the patient when the access covers are open.•