EasyManua.ls Logo

Baylis RFP-100A - WARNINGS, PRECAUTIONS, AND ADVERSE EVENTS; WARNINGS

Default Icon
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© 2012-2019 Baylis Medical Company Inc. 10 of 48 DMR RFP-100A 3.3 V-13 29-May-2020_EN.docx
SECTION 3: WARNINGS, PRECAUTIONS, AND ADVERSE EVENTS
The safe and effective use of RF energy is highly dependent upon factors under the control
of the operator. There is no substitute for properly trained operating room staff. It is
important that the operating instructions supplied with the Generator be read and understood
before use.
3.1. WARNINGS
DO NOT attempt to operate the Generator before thoroughly reading this User’s Manual.
It is vital that the operating instructions for the equipment be read, understood, and
followed properly. For future reference, retain this User’s Manual in a convenient, readily
accessible place.
The Generator is intended for use with separately cleared BMC RF Devices, BMC
connector cables, and the accessory footswitch only. For respective devices/accessories,
refer to individual IFUs for more information.
To avoid risk of electric shock, Generator must only be connected to supply mains with
protective earth.
Do not remove the cover of the Generator. Removal of the cover may result in injury and/or
damage to the Generator.
When the Generator is activated, conducted and radiated electrical fields may interfere with
other medical and electrical equipment. Care should be taken to limit the effects that
electromagnetic interference (EMI) produced by the Generator has on other equipment.
Laboratory staff and patients can undergo significant x-ray exposure during RF Puncture
procedures due to the continuous usage of fluoroscopic imaging. This exposure can result
in acute radiation injury as well as increased risk for somatic and genetic effects.
Therefore, adequate measures must be taken to minimize this exposure.
Do not attempt to perform an RF puncture with an initial cut setting other than that
recommended by the BMC RF Device Instructions for Use. The cut setting (and therefore
output power) should be as low as possible (as recommended for BMC RF device) to avoid
any unintended result.
Failure of the Generator could result in an unintended increase of output power.
Place monitoring electrodes as far away from the surgical site as possible, to avoid burns
or interference with other equipment. The use of needle monitoring electrodes (or other
small area electrodes) during RF output is not recommended. In all cases, incorporating
high frequency current limiting devices are recommended.

Table of Contents