© 2012-2019 Baylis Medical Company Inc. 28 of 48 DMR RFP-100A 3.3 V-13 29-May-2020_EN.docx
7.6. TURN THE GENERATOR “ON”
Turn the Generator “ON” by pressing the rocker switch located on the Generator back
panel to the “I” position. The Generator immediately performs a self-test of power
generation, measurement, and control circuitry, as indicated by the progress bar on the
display (POST state). If there are no errors, the Generator enters into the STANDBY state.
If the self-test fails, a tone is sounded and the Generator enters into the FAULT state, where
the red FAULT INDICATOR flashes and an error code is displayed.
If the FAULT state is entered (i.e. a system malfunction is detected during self-test), the
Generator will not operate. To clear any malfunctions found during the self-test, the
Generator power must be cycled (powered “OFF” and then back “ON”) with the self-test
repeated. If the Generator fails again, the Generator will not function properly and the
error code should be recorded. Baylis Medical Company should be contacted for service.
The Generator will NOT operate unless the power on self-tests have been successfully
completed.
If the dispersive electrode is disconnected or a dual foil dispersive electrode is not
adequately applied, the Return Electrode Fault Indicator will illuminate red.
7.7. SET CUT MODE
Set the desired CUT mode as recommended by the BMC RF Device Instructions for Use,
by using the right soft keys to increment ▲ or decrement ▼ through the settings.
Note: The CUT ▲/▼ soft keys are inoperative during RF power delivery.
7.8. SET TIME
Set the desired duration (in seconds) for RF power delivery, as recommended by the BMC
RF Device Instructions for Use, by using the left soft key to increase ▲ or decrease ▼ the
setting.
Note: The TIME ▲/▼ soft keys are inoperative during RF power delivery.
7.9. CONFIRM GENERATOR SETTINGS AND STATE
Prior to delivering RF energy to the BMC RF Device, first check that all the connections
have been made properly and that the TIME and CUT settings are correct. The Generator
should be in READY state if all connections have been properly made. Confirm that all of
the requirements specified in the individual accessories’ Instructions for Use have been
met. Only after all the above conditions are met should one proceed to deliver RF energy.
7.10. ACTIVATE RF ENERGY DELIVERY