© 2012-2019 Baylis Medical Company Inc. 29 of 48 DMR RFP-100A 3.3 V-13 29-May-2020_EN.docx
RF energy delivery is activated when the RF ON/OFF BUTTON is pressed once or when
the footswitch is pressed and held. Pressing the RF ON/OFF BUTTON again or releasing
the footswitch before the timer elapses will terminate RF energy delivery. When RF power
is delivered to the device, the Generator enters the ON state.
If interference with other equipment is suspected, reposition all cables being sure to keep
Generator cables away from monitor equipment cables.
7.11. DEACTIVATE RF POWER DELIVERY
RF energy delivery is terminated and the READY state is entered when either the timer
elapses, when the RF ON/OFF BUTTON is pressed, or when the FOOTSWITCH is
released during the ON state. The RF ON/OFF BUTTON will extinguish and the tone will
no longer sound.
RF energy is also terminated with an ALERT or with an ERROR. If an ALERT terminated
energy delivery, the alert message is displayed for five (5) seconds and the Generator enters
READY state. If an ERROR terminated energy delivery, the error code is displayed, the
FAULT Indicator flashes, and the Generator enters FAULT state. To attempt to exit the
FAULT state, the Generator power must be cycled, after which the sequence of initializing
steps must be performed again from Section 7.7.
Note: If an ERROR is encountered repeatedly, the Generator is not functioning properly
and needs servicing; Contact Baylis Medical Company.
7.12. RE-APPLY RF POWER
To re-apply RF energy, repeat Steps 7.7 through 7.11. Confirm that the dispersive
electrode is properly applied and connected before adjusting to a higher CUT setting.
7.13. PROCEDURE COMPLETE
When the Generator is no longer required, turn the Generator “OFF” by pressing the
rocker switch located on the Generator back panel to the “O” position.
The BMC RF Device should be disconnected from the BMC connector cable. The BMC
connector cable and return electrode should be disconnected from the generator front
panel. The BMC RF Device, BMC connector cable and return electrode should be
disposed of or stored as indicated in their instructions for use or according to the
procedures for the institution.
If other connectors were used (e.g. footswitch, USB, etc), they should be disconnected as
necessary to ensure safe storage of the Generator and accessories.