SECTION 1: DEVICE DESCRIPTION
The Generator is a component of the Baylis Medical Company Radiofrequency (RF) Puncture
System. The Generator is operated in conjunction with separately cleared BMC RF Devices, BMC
connector cables such as (but not limited to) the RFX-BAY-TS, RFX-BAY-TS-12-SU or RFX-
BAY-OTW-10-SU cables, an optional DuoMode
TM
Cable, a commercially available return
(dispersive) electrode that meets or exceeds IEC 60601-2-2:2017, and an optional accessory
footswitch. The Generator delivers energy in a voltage-controlled monopolar mode between the
BMC RF Device’s distal tip electrode and the return electrode. Detailed information regarding the
BMC RF Device is contained in a separate manual that accompanies each BMC RF Device.
The Generator produces continuous radiofrequency power output at a fixed frequency in the range
of 450 kHz to 480 kHz in a monopolar mode. Connections for the Generator connector cable
(which connects to the separately approved BMC RF Device) and a patient return electrode that
meets or exceeds IEC 60601-2-2:2017 are provided. Controls on the front panel allow the cut mode
and the duration of the RF output to be set. In addition, on/off control of the output can be achieved
through the optional accessory footswitch or through the dedicated front panel button. The elapsed
time and the cut mode are displayed on the liquid crystal display (LCD) during RF energy delivery.
An audible tone synchronized with the RF output is also produced during energy delivery. The
Generator has several built-in safety features, such as device identification, alert messages, an
automatic shut-off for out-of-range parameters or metal contact, and maximum voltage, current,
and power limits.
The Generator has been tested for compliance with the following standards:
IEC 60601-1:2005+A1:2012
IEC 60601-2-2:2017
IEC 60601-1-2:2014