EasyManua.ls Logo

Baylis RFP-100A - CONTROLS, DISPLAYS, AND CONNECTIONS; FRONT PANEL DISPLAYS, CONTROLS, AND CONNECTIONS

Default Icon
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© 2012-2019 Baylis Medical Company Inc. 15 of 48 DMR RFP-100A 3.3 V-13 29-May-2020_EN.docx
SECTION 5: CONTROLS, DISPLAYS, AND CONNECTIONS
5.1. FRONT PANEL DISPLAYS, CONTROLS, AND CONNECTIONS
Descriptions of the front panel displays, controls, and connections are given below. Refer to
Figure 5-1 - Generator Front Panel for their location.
(1) AC Power Indicator: This green LED illuminates when the Generator is turned on.
(2) FAULT Indicator: This red LED illuminates and flashes when a system ERROR has
occurred. System errors include self-test failures, hardware protection errors, hardware
measurement errors, and software failures. Main power to the Generator must be cycled
(off-on) to attempt recovery from a system error. Consult instructions for use.
(3) Ambient Light Sensor: This sensor detects the ambient light level. Screen brightness is
automatically adjusted according to ambient light level (bright (HIGH) in a bright room
and dimmed (LOW) in a dim room).
(4) Return Electrode Fault Indicator: The red LED illuminates when a return electrode is
NOT connected to the generator OR when the measured impedance of a monitoring (dual
foil) return electrode is greater than 150 ohm, indicating poor patient contact. Note: Only
use return electrodes that meet or exceed IEC 60601-2-2:2017 requirements.
8
3
9
5
1
2
6
7
14
12
4
10
11
Figure 5-1 - Generator Front Panel
15

Table of Contents