© 2012-2019 Baylis Medical Company Inc. 36 of 48 DMR RFP-100A 3.3 V-13 29-May-2020_EN.docx
9.4. ALERT CODES
Summary of Alert Codes, Displayed Text, and Description
High Impedance Detected.
Check device and all cable
connections
Greater than 6000 ohm measured during
RF energy delivery. May indicate poor
connection between Device and connector
cable or poor patient contact with
dispersive electrode.
Less than 100 ohm measured during RF
energy delivery. Device may be in contact
with a metal object.
Check ground pad contact
with patient. Replace
grounding pad if necessary.
Impedance between the conductors of a
dual foil dispersive electrode is greater
than 150 ohm. This may indicate poor
patient contact.
Check ground pad contact
with patient. Replace
grounding pad if necessary.
Impedance between the conductors of a
dual foil dispersive electrode is greater
than 150 ohm. This may indicate poor
patient contact.
Check ground pad contact
with patient. Replace
grounding pad if necessary.
Impedance between the conductors of a
dual foil dispersive electrode is greater
than 150 ohm. This may indicate poor
patient contact.
Check all ground pad
connections and ground pad
contact with patient.
Open circuit between conductors of
dispersive electrode connection. May
indicate faulty dispersive electrode or
poorly connected dispersive electrode.
Error with connector cable may have
altered Device specific settings. The
connector cable should be checked for any
problems.
Invalid device detected.
Record alert code and contact
Baylis Clinical Support.
Invalid or broken connector cable
connected to the Generator.
Unsupported device detected.
Record alert code and contact
Baylis Clinical Support.
Device does not have supported output
parameters.
Connector cable was disconnected or
failed during RF energy delivery.
The connected device
conflicts with the current
channel configuration.
Device parameters have been corrupted.
Contact Baylis Clinical Support.
Either connector cable is not functioning
or not connected when attempting to