© 2012-2019 Baylis Medical Company Inc. 37 of 48 DMR RFP-100A 3.3 V-13 29-May-2020_EN.docx
Summary of Alert Codes, Displayed Text, and Description
Alert Code
Displayed Text
Description
deliver RF energy. Check connector cable
connection, replace connector cable if
necessary.
A019
Metal contact was detected.
Reposition the device.
Metal contact detection feature has
terminated RF energy delivery due to
device proximity to metal. Reposition
device before attempting to deliver RF
energy.
A020
The treatment time limit has
been exceeded.
RF delivery longer than Time setting. If
persists, contact Baylis Clinical Support.
A021
The device is not ready for
treatment.
RF ON/OFF button has been damaged or
is stuck. If persists, contact Baylis Clinical
Support.
A022
The device is not ready for
treatment.
Footswitch has been damaged or is stuck.
Disconnect footswitch and deliver RF
using RF ON/OFF button. Csontact Baylis
Clinical Support.
A023
Treatment was terminated
early.
RF delivery shorter than Time setting. If
persists, contact Baylis Clinical Support.
A024
Settings have changed
according to device limits.
Time and Cut setting in Standby state were
either out of range for a given Device or
the Time setting was out of range for the
desired Cut mode. Settings are
automatically adjusted to maximum valid
setting.
A025
Settings have changed
according to device limits.
Time and Cut setting in Standby state were
either out of range for a given Device or
the Time setting was out of range for the
desired Cut mode. Settings are
automatically adjusted to maximum valid
setting.
A026
Record Alert Code and
contact Baylis Clinical
Support.
Generator Setup settings have been lost.
Settings have been reset to factory
defaults.
A027
An unknown failure occurred
while exporting treatment
data.
Export of treatment data failed. Check
USB flash disk connection. If persists,
contact Baylis Clinical Support.
A028
Not enough space free to
export treatment data.
USB flash disk is full. Free up space
before attempting to obtain treatment logs.
A029
Invalid device detected.
Record alert code and contact
Baylis Clinical Support.
Invalid connector cable connected to the
Generator. Replace connector cable.