© 2012-2019 Baylis Medical Company Inc. 20 of 48 DMR RFP-100A 3.3 V-13 29-May-2020_EN.docx
SECTION 6: DISPLAYS
6.1. System Initialization and POST States
The System Initialization state is initiated when the Generator is turned on. It lasts for
approximately 40s.
The AC Power Indicator is illuminated and a splash screen appears. The screen will
go blank for approximately 20 seconds.
Figure 6-1- System Initialization Display
The Power On Self Test (POST) state is initiated after System Initialization is
complete. It lasts for approximately 10s.
The FAULT indicator illuminates during POST.
The Return Eletrode Fault Indicator and RF ON/OFF indicator flash briefly during
POST.
A tone is sounded once POST is successfully completed.
If the Generator does not enter POST, then it will be considered as an error (see
FAULT state).
Figure 6-2- POST Display