EasyManua.ls Logo

Baylis RFP-100A - CONNECT THE DISPERSIVE (RETURN) ELECTRODE

Default Icon
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© 2012-2019 Baylis Medical Company Inc. 27 of 48 DMR RFP-100A 3.3 V-13 29-May-2020_EN.docx
Refer to the Connector Cable Instructions for Use for further details.
To connect the BMC RF Device to the BMC Connector Cable, please refer to the individual
Instructions for Use of each BMC RF Device.
Note: Position the Generator near the table where the procedure is to be performed. The
Generator may only be connected to recording systems providing patient electrical
isolation in accordance with IEC 60601.
7.5. CONNECT THE DISPERSIVE (RETURN) ELECTRODE
Only use dispersive electrodes that meet or exceed IEC 60601-2-2:2017 requirements.
Connect the dispersive (return) electrode (also referred to as a grounding pad) connector to
the Return Electrode Connection found on the Generator front panel. Line up the pins in
the Generator receptacle with the sockets of the electrode connector and gently push the
dispersive electrode connector until it sits firmly in place. To unplug, grasp the dispersive
electrode connector and gently pull it out from the receptacle.
Before use, it is important to check the dispersive electrode’s sealed foil packaging for
damage. Exposure to air, due to a damaged package, could cause the dispersive electrode
to become dry and limit its capability to provide an appropriate return path for RF energy.
Be sure the pad is moist and sticky to the touch before placing it on the patient. Do not
attempt to relocate the patient dispersive electrode after its initial application. Electrode
gel that is separate from and additional to any gel that is an integral part of the dispersive
electrode is not required and should not be used.
The dispersive electrode should be placed on a well-vascularized, convex skin surface.
Dispersive electrode placement on the thigh may be associated with higher impedance.
Avoid scar tissue, bony prominence, adipose tissue, or any areas where fluid may pool.
Shave, clean, and dry the application site as needed. Check for wrinkles or folds when
applying the dispersive electrode, as they can inhibit adequate contact and decrease
conductivity.
The Generator is equipped with a Return Electrode Fault Monitor that measures the
impedance between the two conductors of the dispersive electrode. For a single foil
electrode, the monitor will indicate a fault (the Return Electrode Fault Indicator will
illuminate red) if either one or both of the conductors are faulty. For a dual foil electrode,
the monitor will indicate a fault if the impedance between the conductors is greater than
150 ohms, which may indicate poor patient contact.
WARNING: Unless a compatible dual foil dispersive electrode is used that meets or
exceeds IEC 60601-2-2:2017 requirements (such as ConMed
®
MacroLyte
®
Dual
Dispersive Electrode), loss of safe contact between the dispersive electrode and the patient
may not result in an auditory alarm.

Table of Contents