EasyManua.ls Logo

Baylis RFP-100A - STANDBY State; READY State; Figure 6-3- STANDBY State Display; Figure 6-4 A- READY State Display

Default Icon
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© 2012-2019 Baylis Medical Company Inc. 21 of 48 DMR RFP-100A 3.3 V-13 29-May-2020_EN.docx
6.2. STANDBY State
The Standby State is initiated upon successful completion of the POST state.
TIME and CUT settings can be adjusted with the left and right soft keys beside the
arrows.
Messages instruct the user to connect a Valid Device and Return Electrode
(Grounding Pad).
RF energy delivery cannot be initiated.
Figure 6-3- STANDBY State Display
6.3. READY State
The Ready State is initiated when a Return Electrode (Grounding Pad) is connected
AND a valid device is connected OR when RF energy delivery is terminated.
TIME and CUT settings can be adjusted with the left and right soft keys beside the
arrows.
RF energy delivery can be initiated by either pressing the RF ON/OFF button or
pressing and holding FOOTSWITCH.
Figure 6-4a- READY State Display

Table of Contents