SECTION 7: DIRECTIONS FOR USE
7.1. READ INSTRUCTIONS FOR USE
Do not operate the Generator or BMC RF Device before thoroughly reading their
respective Instructions for Use. It is vital that the Instructions for Use for all the associated
equipment is read, understood, and followed properly.
7.2. CONNECT GENERATOR POWER CORD
Connect the Generator Power Cord plug into a properly grounded AC electrical outlet. To
ensure proper grounding, the Power Cord plug must be installed in an AC electrical wall
outlet designated “Hospital Grade” or “Hospital Only”. Never use an outlet without a
grounding connection.
Position the Generator for easy viewing of its front panel displays.
7.3. CONNECT FOOTSWITCH (OPTIONAL)
If the footswitch is to be used with Generator, it must be connected to the Generator rear
panel. Line up the four (4) pin footswitch connector with the footswitch connection. Rotate
the connector until it slides and clicks into place. A light tug on the cable can confirm that
the connector is properly seated.
To disconnect the footswitch from the Generator, grasp the connector housing and gently
pull it straight out of the receptacle. The connector housing slides backwards to disengage
the locking mechanism.
Never disconnect the footswitch by pulling on the cable cord.
7.4. CONNECT CONNECTOR CABLE
Verify the BMC Connector Cable model specified by the BMC RF Device Instructions for
Use. Connect the four (4) pin connector to the Connector Cable Connection on the
Generator front panel. The connector on the Connector Cable that connects to the
Generator is keyed. Gently line up the key while pushing in gently until the connector
clicks firmly into the receptacle.
To disconnect the Connector Cable from the Generator, grasp the connector housing and
gently pull it straight out of the receptacle. The connector housing slides backwards to
disengage the locking mechanism.
Never disconnect the Connector Cable by pulling on the cable cord.
Do not twist the connector of the BMC Connector Cable while inserting or removing it
from the Generator receptacle. Doing so may damage the pin connections.

Table of Contents