© 2012-2019 Baylis Medical Company Inc. 16 of 48 DMR RFP-100A 3.3 V-13 29-May-2020_EN.docx
(5) STATE Status Bar: This window displays the current STATE of the Generator. The
various Generator states and how they are related are shown in Figure 5-3 – Generator
States Flow Chart.
(6) TIME Setting Window: This window displays the desired RF output duration (in
seconds). This window also displays the functions for the soft keys used to adjust the
TIME setting.
(7) Left column soft keys: These keys allow for parameter adjustment. Their function is
displayed on the screen to the right of the soft key, when applicable. For example, the up
▲ and down ▼ arrows are displayed when the keys are to be used to increase or decrease
a setting.
(8) CUT Setting Window: This window displays the desired CUT mode. CUT modes are
BMC RF Device specific and described in Section 9.3. The BMC RF Device’s
instructions-for-use should be consulted for appropriate settings to use. This window also
displays the function for the soft keys used to adjust the CUT setting.
(9) Right column soft keys: These keys allow for parameter adjustment. Their function is
displayed on the screen to the left of the soft key, when applicable. For example, the up
▲ and down ▼ arrows are displayed when the keys are to be used to increase or decrease
a setting.
(10) Bottom row soft keys: These soft keys have various functions. Each key’s function is
displayed on the screen above the soft key when applicable.
(11) Message Window: This window displays functional and informational messages when
required.
(12) Return Electrode Connection: This patient isolated connection is for attachment of an
approved dispersive (return) electrode. Only use dispersive electrodes that meet or
exceed IEC 60601-2-2:2017 requirements. Either non-monitoring (single foil) or
monitoring (dual foil) electrodes can be used.
(13) Connector Cable Connection: This patient isolated connection is for the attachment of
the RFP-100A connector cable. The user shall refer to the BMC RF Device instructions-
for-use to select the proper connector cable model.
(14) RF ON/OFF Button and Indicator: Upon press and release, this button initiates RF
energy delivery when the Generator is in the READY state. This button terminates RF
energy delivery when the Generator is in the ON state. The indicator in the button
illuminates yellow when the Generator is in the ON state.
(15) USB Port (side): When a USB memory stick is connected, treatment data from the last
twenty (20) RF energy deliveries is downloaded. When left connected, data from
subsequent RF deliveries is downloaded on a per treatment basis.