Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
Exterior surfaces,
inner chamber,
door gaskets
Standard commercial cleaning detergents free from acid or halogenides.
Alcoholic solutions.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Expansion racks,
rack holders
Standard commercial cleaning detergents free from acid or halogenides,
no salt solution or chlorinated solvents.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Instrument panel Standard commercial cleaning detergents free from acid or halogenides.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
For surface protection, perform cleaning as quickly as possible.
After cleaning completely remove cleaning agents from the surfaces with a moistened towel.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.
CAUTION
Danger of corrosion.
Damage to the unit.
∅ Do NOT use cleaning detergents with acids or halogenides.
We recommend using the neutral cleaning agent Art. No. 1002-0016 for a thorough and mild
cleaning.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
Decontamination
Vacuum drying ovens VD can be hot air sterilized at 190 °C / 374°F for at least 30 minutes. Remove any
inflammable goods from the interior before.
With every decontamination method, always use adequate personal safety controls.
17.3 Sending the unit back to BINDER GmbH
If you return a BINDER product to us for repair or any other reason, we will only accept the product upon
presentation of an authorization number that has previously been issued to you. An authorization number
will be issued after receiving your complaint either in writing or by telephone prior to your sending the
BINDER product back to us. The authorization number will be issued following receipt of the information
below:
• BINDER product type and serial number
• Date of purchase
• Name and address of the dealer from which you bought the BINDER product
• Exact description of the defect or fault
• Complete address, contact person and availability of that person
• Exact location of the BINDER product in your facility
• A contamination clearance certificate (chap. 21) must be faxed in advance