When transmitting multiple sets of data to PC after saving, a network error could occur
which may cause transmission halt and subsequent loss of data. Therefore, we
recommend users to practice following procedures to minimize such loss.
1. In EKG Main, press [NET] button to transmit data to PC immediately after
measuring.
2. When transmitting multiple sets of measurements, rather than sending all data at
once, we recommend users to transmit individual records after each
measurement.
If the system boots up without having LAN cable connected, the network function may
not work properly. In this case, enter [System] -> [Network] -> [OK] then press ‘NETWORK’
button to attempt running the network function again. Please refer Network Setup part of
the manual to get more information regarding IP configuration setup.