V 1.31
2021. 05 161 / 171 Doc. # : BN-OP-C7
Transmission failed
file is existing.
Are you sure to
delete?
When the transmission
failed file to PACS store
server needs to be
deleted
Delete the failed file after
transmitting the file to PACS store
server
When the battery level is
Low
Please connect power cable or
charge battery
Printer stops working due
to technical problem of
device
When the wrong printer
data were input
Printer:
Temperature high
When the Printer gets
heated due to long and
continuous use of printer
Please stop the use of printer for
at least one minute and try again
When the printing paper is
completely out
Please put new roll of printing
paper for use
Printer: Printer not
ready
When there is no ‘ready’
answer from printer
module
When the wrong protocol
was put into the printer
When the printer was
abnormally forced to stop
during printing
When the lead is off from
the patient body
Please reattach the electrode on
patient body
When the conductivity
between patient body
and electrode is weak
Apply some water to reusable
electrodes, and reattach them
When the user tries to
execute copy printing
without any recording
data
Try make copy after recording
data by “RECORD”, “AUTO”, and
“NET” button