26 | 405 TS Getting Started Guide
Verify the Washer
1. Fill the washer’s supply bottle (bottle “A” for buffer switching models) with
approximately one liter of deionized water.
LHC Touch screen
1. Select File > Open.
2. Open the 405 TS and the Prime and
Maintenance Protocols folders.
Open the applicable "Buffer" folder,
and then open W-DAY_RINSE.LHC
You may need to reset the Com
Port.
3. When ready, click the Run button
to prime the tubing and manifold
with deionized water.
4. When finished, close the program.
1. Select Maintenance at the Home
screen.
2. Select W-Day_Rinse and press
Start.
3. When the protocol is completed,
press the Home button.
Run the Verify Test ("Q" Models Only)
If your 405 TS is equipped with BioTek's Verify™ Technology, run the test:
Important! Before running the test:
Make sure the 405 TS is primed and ready to run:
l Fully prime the tubing/system, e.g. run W-Day Rinse.
l If necessary, empty waste vessel and tighten waste bottle cap.
Prepare to run the Verify test:
l Fill the supply vessel with at least 100 mL dH2O or DIwater or buffer solution
1
.
Run the Verify test:
Touch screen
1. Select Quick>Verify Manifold and specify the parameters:
1
Do not use highly viscous fluids or wash buffers that are prone to leave significant residue on the plate.
405™ TS Microplate Washer