Verify Performance| 27
l Buffer: Select the bottle to use, if applicable.
l Prime: optionally, prime the manifold tubes to correct for evaporation loss. 40
mL of fluid is dispensed.
l %CV Threshold: optionally, change the %CV to match your lab's standard for
QC tests: 5-15%. The default value of 5% is recommended.
2. Press Start to run the test.
3. Assess the results: Verify Results.
Do not discard the Verify Test Plate. Reuse it for as long as possible.
405 Touch: BioTek recommends exporting the initial test results and saving them for
future reference.
Contact BioTek TAC if any of the tests fail.
BioTek Instruments, Inc.