Verify Performance| 35
Frequency:
5 Minutes: Gain complete confidence that your plates will be processed correctly by
spending five minutes to run the Verify test before running your assay!
Develop a schedule for running the Verify routine based on the type of wash buffer
you are using, how frequently you use the washer, and your maintenance habits:
l Daily: When using PBS, protein, and other fluids that easily crystallize and
harden, potentially clogging the tubes, run the Verify routine every day and/or
after prolonged downtimes to ensure the tubes are not clogged.
l Weekly: When using TRIS and other low-salt buffers and when adhering to the
recommended maintenance schedule, you can confidently run the Verify routine
less frequently, e.g. before new assay runs.
Replace the Verify Test Plate (according to the prescribed procedure) if it is
damaged, scratched, warped, etc.
For more details: See About Verify Technology in the operator's manual.
Handling the Verify Test Plate
The Verify Test Plate ships in its own special box. The plate is labeled with the serial
number of the 405 TS that has been calibrated for its use.
l Well A1 is opposite the
chamfered or beveled
edges.
l The plate is labeled with
the washer's serial
number.
l When the plate is on the
carrier in the proper
orientation, the "Front"
label is visible.
Verify Test Plate