(-221) Settings conflict
(-222) Data out of range
(-223) Too much data
(-224) Illegal parameter value
(-225) Out of memory
(-230) Data Corrupt or Stale
(-310) System error
(-350) Too many errors
(-400) Query error
(-410) Query INTERRUPTED
(-420) Query UNTERMINATED
(-430) Query DEADLOCKED
(223) Front panel buffer overrun
(224) Front panel timeout
(225) Front Crc Check error
(401) CAL switch prevents
(402) CAL password is incorrect
(403) CAL not enabled
(404) Readback cal are incorrect
(405) Programming cal are incorrect
SYSTem:VERSion?
This command is used to query the current SCPI version. The returned parameter is a string
like”YYYY.V”, in which the YYYY is the year of that version, V is the software version of that year.
Command syntax: SYST:VERS?
Parameter: None
Returned parameter: <NRf>
SYSTem:REMote
This command is used to switch to the remote control mode (PC control). This command is used
to set the power supply to remote control mode through the RS232 interface. Except for the
Local key on the front panel, other keys are locked and cannot be used. Sending or receiving
commands without first sending the command to configure the remote control may cause
unpredictable results.
Command syntax: SYST:REM
Parameter: None
SYSTem:LOCal
This command is used to switch the instrument to local control mode (front panel control).
Command syntax: SYST:LOCal
Parameter: None
SYSTem:RWLock
This command is the same function as SYSTem:REMote, except this command can lock “LOCAL”