13
P073GUI001 - e-Celsius Medical Connect System User Manual.4 - EN
The system is based on the principles listed in Table 3.
Table 3: Operating principle
N° Operating principle
1
The e-Celsius® Medical capsule is activated by an electromagnetic pulse emitted by the
activator
2
The e-Celsius® Medical capsule measures temperature using a Silicon Bandgap temperature
sensor
3 The e-Celsius® Medical capsule automatically stores the last 2,000 collected data
4
The communication between the e-Celsius® Medical capsule and the e-Med Connect watch is
realized by radio frequency on the 433MHz band via a proprietary protocol
5 The e-Med Connect watch receives and stores real time data at a period of 30s
6
The e-Med Connect watch automatically re-requests any unreceived data from the e-Celsius®
Medical capsule in real time
7 Viewing data on the e-Celsius® Mobile application
8 Data export to CSV and PDF formats via the e-Celsius® Mobile application
9 Physiological alarm function for data exceeding high or low temperature thresholds.
Frequently used functions are listed in Table 4.
Frequently used funcons
Primary (P) /
Secondary (S)
Installaon of the e-Celsius Mobile applicaon on the smartphone or Android tablet P
Turning on the e-Med Connect watch P
Creaon of a prole to associate the watch data with this prole P
Pairing of the e-Med Connect watch by BLE with the support of e-Celsius Mobile (smart-
phone or tablet)
P
Seng the me and date of the e-Med Connect watch via the e-Celsius applicaon®
Mobile
P
Seng of alarms (thresholds) by the e-Celsius soware® Mobile then transmied to the
e-Med connect watch
P
Seng the RF channel used by the e-Med Connect watch S
Acvaon of the e-Celsius capsule® Medical P
Ingeson of the e-Celsius Medical capsule with a glass of water P
Temperature measurement with the e-Celsius capsule® Medical every 30s. P
Storage of the last 2000 data in the e-Celsius capsule® Medical P
Automac transfer of real me data at a period of 30s by 433MHz RF from the e-Celsius
Medical capsule to the e-Med Connect watch
P
Automac synchronizaon of data from the e-Celsius® Medical capsule by the e-Med
Connect watch of the last 2000 data
P