33
P073GUI001 - e-Celsius Medical Connect System User Manual.4 - EN
When a capsule reaches the end of its life, an alarm is triggered on the watch (ashing LED) and the
information is then transmitted to the e-Celsius Mobile application as a notication.
The capsule will stop about 500 measurements after the notication appears. A marker is automatically
placed in the data le.
In order to reset the alarms and the Min/Max values shown in the data display window, click and hold
your nger on the alarm symbol. A window will appear offering you a "Reset alarms". This action will then
concern all the capsules associated with the watch. A conrmation message will be asked to validate the
reset.
When the use is nished and you want to stop the capsule, to do so, simply go to
the e-Celsius Mobile application, on the concerned prole and choose the capsule
you want to stop.
Hold your nger on the cap and slide it to the right.
In order to ensure the success of the procedure, the capsule and the watch must
be close enough to communicate.
A conrmation request must be validated before the actual
disassociation of the capsule. This action is denitive, the
capsule disappears from the database of the watch. The data
le corresponding to the capsule is archived in the application
e-Celsius Mobile.
* the display of the °C! symbol at the top left of the watch's display
* the symbol on the prole and on the concerned capsule
* the min or max associated with the cap in Red or Blue color
* a notication on the phone/tablet if it is connected in real time BLE to the watch
The delay inherent in the determination of an alarm condition is a maximum of 30 seconds in the case of
real-time communication between the capsule and the watch.
This physiological alarm has a lock. These different signals will remain visible until the alarms are reset.
When the current temperature stabilizes between the two thresholds, the alarm signals remain visible until
the alarms are reset.
The alarm system is divided between the watch and the mobile application. The two elements of the
system must allow a BLE communication. For this, they must be within 10m of each other and the watch
worn on the patient's wrist.
Note that the alarm may be triggered by synchronized data (received afterwards) that has previously
exceeded one of the two temperature thresholds. In this case, the minimum or maximum values are time-
stamped at the time the data was received (date of synchronization).
VISUALIZATION OF THE END OF LIFE OF THE CAPSULE
DESASSOCIATION OF THE CAPSULE
RESET ALARMS
Figure 15 - capsule disassociation