11 Technical specifications
900-350_IFU_V2.6_10296-S13-20100111-EN ARC 300 / ARC 350 Operating manual 85
Setting Program
CardiacCUT 14, 15
ARC350CCUT_SE_080124_1
0
10
20
30
40
50
60
70
80
90
100
10 100 1000 10000
P [W]
CardiacCut
90W
CardiacCut
45W
Measurement of resistances
• Diagram of power output P [W] as a
function of the load resistance R [Ω] for
the setting “CardiacCUT” = 90 W/45 W
ARC350CCUT_SE_080124_2
0
10
20
30
40
50
70
60
80
90
0102030405060708090
displayed value
P [W]
• Diagram of power output P [W] as a
function of the “CardiacCUT” setting.
Measured load resistance = 500 Ω
ARC350CCUT_SE_080124_3
0
500
1000
1500
2000
2500
3000
0
displayed value
U [Vp]
10 20
30 40
6050 70 80 90
• Diagram of HF output voltage U [Vp] as
a function of the "CardiacCUT" setting
(idle mode)