Page #43 background image
43 How to Use the Clearway 2
Clearway 2 User Manual
DOC-001000 Ver 6
Entraining Oxygen
Please ensure you read the "The Usage of Oxygen if Prescribed” section (page
18) before using oxygen with the Clearway 2.
Oxygen is a prescribed medicine and must only be authorised
by a competent clinician.
Always referrer to the oxygen provider’s instructions for safety.
It is possible to use up to 15 litres per minute of low flow oxygen with the
Clearway 2 if prescribed by your referring clinician. Only use the amount of
oxygen as advised.
Never turn the oxygen on before turning the device on, as
oxygen may pool around the machine.
To entrain Oxygen into the device safely, connect the oxygen into the circuit
as close to the patient as possible using an oxygen connector (see Accessories, page
78).
Once the treatment has finished, ensure the oxygen is turned off,
before turning off the device.
Use with SpO
2
Sensor
The SpO
2
displays the patient’s oxygen saturation, if measured with SpO
2
module accessory. When using the Clearway 2 with the SpO
2
sensor accessory
(see Accessories, page 78) it is calibrated to display functional oxygen saturation.
To use the SpO
2
sensor with the Clearway 2, it must be connected to the side
of the device (see Side Panels, section 3.3).

Table of Contents

Other manuals for Breas Clearway 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Clearway 2 and is the answer not in the manual?

Breas Clearway 2 Specifications

General IconGeneral
BrandBreas
ModelClearway 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals