Safety Information 16
DOC-001000 Ver 6 Clearway 2 User Manual .
Clearway 2 should be observed to verify normal operation in that
configuration.
Mobile or transportable radio transmitters may interfere with the
Clearway 2 or performance of the display. Guidance for safe
positioning of the Clearway 2 available on page 76.
If a portable AC power supply is used, make sure that the voltage
variations are within the operating limits of the Clearway 2. See
Electrical Information” on page 68 for AC operating limits.
Use of accessories, transducers and cables other than those
specified
or provided by Breas could result in increased
electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.
Environmental Safety
To ensure environmental safety is maintained ensure the below
procedures are met
Do not use the Clearway 2 in any toxic environment.
Do not use the Clearway 2
in environments where there are
explosive gases or other flammable anaesthetic agents present.
The performance of the Clearway 2
may deteriorate at ambient
temperatures below -4°F (-20°
C) and above 104°F (40°C).
However, the treatment shall always be started in an ambient
temperature warmer than 41°F (5°C).
Do not expose the Clearway 2 to rain or snowfall.
Do not use or store in the presence of strong electromagnetic
fields such as an MRI environment. Use of the Clearway 2 in an
MR environment may result in malfunction of the Clearway 2 and
pose unacceptable risk to the patient, medical staff, or other
persons.

Table of Contents

Other manuals for Breas Clearway 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Clearway 2 and is the answer not in the manual?

Breas Clearway 2 Specifications

General IconGeneral
BrandBreas
ModelClearway 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals