77 Technical Specifications
Clearway 2 User Manual
DOC-001000 Ver 6
Recommended separation distances between external power
conductors and the Clearway 2
Rated Maximum Current in
Conductor (A)
Separation Distance (M)
50-60 Hz
d= I/2πH= I/188
1
0.005
10
0.05
30
0.16
For conductors rated at a maximum current not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation d=I/2πH, where I is the maximum current rating of the conductor
in amperes (A) according to the transmitter manufacturer; H is the Clearway
2 immunity compliance level to electromagnetic fields in the 50-60 Hz
frequency span (30 A/m).
Portable RF devices (e.g. mobile phones) should be used no
closer than 30 cm (12 inches) to any part of the ME Equipment,
including any cables
Clearway 2 in other countries
If you are traveling to a country with a line voltage which is different to where
you were issued the device, then a different power cord or an international plug
adaptor may be required to ensure your power cord is compatible with the
power outlet of the country where you are travelling.

Table of Contents

Other manuals for Breas Clearway 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Clearway 2 and is the answer not in the manual?

Breas Clearway 2 Specifications

General IconGeneral
BrandBreas
ModelClearway 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals