Technical Specifications 68
DOC-001000 Ver 6 Clearway 2 User Manual .
Measurable Flow
0 - 500 l/min
Resolution 1 L/m
Tolerance ± 10%
Breath Rate (NIV)
6 - 60 BPM
Resolution 1 BPM
Tolerance ± 1 BPM
SpO
2
85 - 100%
Tolerance ± 3 digits
No motion or flex sensor
Pulse Rate
18 - 300 BPM
Tolerance ± 3 digits.
No motion or flex sensor
ALARM MEASURED SOUND LEVEL INDICATION
Medium Priority
Audible Alarm
45 - 50 dB(A)
± 4 dB(A)
Measured at 1 m
Low Priority
Audible Alarm
45 - 50 dB(A)
± 4 dB(A)
Measured at 1 m
Electrical Information
ELECTRICAL FEATURE SPECIFICATIONS
AC Voltage Source
100 to 240 V AC, tolerance +10%/-20%,
50 to 60 Hz, max 260 W
Type of protection against
Electric Shock
Class II
Degree of protection
against Electric Shock
Type BF Applied Part
Ingress Protection Rating Ingress Protection rating: IP22
Mode of Operation Continuous