71 Technical Specifications
Clearway 2 User Manual
DOC-001000 Ver 6
Emissions and Immunity Declaration
Note: The ‘Essential Performance’ of the Clearway 2 is defined as follows:
Clearway 2 Essential Performance
The Clearway 2 will deliver therapeutic treatment via the patient-connection
port (outlet) at the set pressure (+/- 10% of full scale).
Under the immunity test conditions of IEC 60601-1-2 4
th
Edition, the
following allowances are acceptable:
If the SpO
2
sensor accessory is connected, the Clearway 2 will provide
SpO
2
readings within its published accuracy specifications or, indicate
abnormal operation. Any temporary degradation of SpO
2
performance
following transient immunity test exposure shall recover from any
disruption within 30 seconds.
Additionally, the following shall not be allowed:
Permanent damage or unrecoverable loss of function
Changes in programmable parameters or settings
Reset to default settings
Change of operating mode
Initiation of unintended operation, or change of operation

Table of Contents

Other manuals for Breas Clearway 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Clearway 2 and is the answer not in the manual?

Breas Clearway 2 Specifications

General IconGeneral
BrandBreas
ModelClearway 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals