EasyManua.ls Logo

Brooks MFM - Page 7

Brooks MFM
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation and Operation Manual
X-TMF-MfS-MFC-eng
PN 541-C-061-AAG
November, 2008
MF Smart Series
7
Section 1: Introduction
MASS FLOW CONTROLLER
Standard features include:
The supplied inlet adapter has a built-in, upstream inlet lter for ow
straightening. Besides reducing the effects of changes in upstream piping,
this also protects the measuring element from occasional debris in the ow
stream. In addition, it eliminates the need for extra (expensive) inlet ltering.
Each MF-series Smart Mass Flow Controller is provided with an adaptive
valve-control algorithm for the maintenance of stable ow control, even
under varying process conditions (pressure/temperature).
Setpoint source selectable: Setpoint signal source for the Mass Flow
Controller can be selected from: (not for Probus)
communication (RS232 / RS485) or
0-5 Vdc or
1-5 Vdc or
0-20 mA or
4-20 mA.
Fast Response Control provides rapid gas settling times with minimal over/
undershoot. Standard factory setting: 1 sec. for MF50/51 and meter models,
3 sec. for MF53.
Valve Override (VOR, available at pin 12) allows the user to fully open or
close the control valve independently of the set-point.
Low ow cutoff prevents the valve from opening whenever the Set point is
less than 1% of full scale.
ALL MODELS
Digital communication features
Digital communication, via RS485 or RS232, provides access to all of the
Smart TMF’s functions, including:
Accurate mass ow measurement and Set point regulation (controller only),
as a percentage of full scale and in selectable engineering units
Flow totalizer
Ambient temperature
Operational settings
Calibration (storage of up to 10 calibration curves)
Controller settings (controller only)
- PID settings
- Selectable soft start
- Adaptive valve control
Adaptive ltering for signal ow component
Alarms
Self-diagnostic
EEPROM error
 - Database error
 - Analogue output error
Out-of-range indications for
- Analogue set points higher than 100% (controller only)
 - Flow
 - Valve (controller only)
 - Analogue output

Table of Contents

Related product manuals