142
300: Block_Logical_ Transfer
0,
16384 -
outpl:JtS
the DMA
block
from
the main mem-
ory. For details
of
what the DMA
block
entails, refer
to
the 2032 main IEC
Interface Manual.
301: Block_Logical
Transfer 17, 2048 - outputs the Time Function in Ch.A,
or
in Ch.B
Spectrum Averaging, the Time Function in Ch.B.
302: Block_Logical_ Transfer 18, 2048 -
outputs
the Time Function in Ch.B.
303: Block_Logical_ Transfer 19, 2048 - outputs the
Autospectrum
Ch.A,
or
in Ch.B
Spectrum Averaging, the
Autospectrum
in Ch.B.
304: Block_Logical_ Transfer 20, 2048 - outputs the
Autospectrum
Ch.B.
305: Block_Logical_ Transfer 21, 4096 -
outputs
the
Cross
Spectrum or, when Signal
·
Enhancement mode is in operation, the Enhanced Time Function Ch.B.
306: Block_Logical_ Transfer 19, 4096 -
outputs
the Enhanced Time Function Ch.A.
307: Block_Logical_ Transfer 23, 2048 - outputs the
A-weighting
function
for
the in-
put buffer, see the
2032 main IEC Interface Manual
for
details.
308: Block_Logical_ Transfer 30, 2048 -
outputs
the
A-weighting
function
for
the
stored
buffer, see the 2032 main IEC Interface Manual
for
details.
1000*:
Dump_Piot-
means that the DIGITAL
1/0
"Output"
pushkey has the same
ac-
tion as the DIGITAL
1/0
"Plot"
pushkey.
1001
*: causes output
of
the 2032 Main
Cursor
reading
1
002*: causes output
of
the 2032
Auxiliary
Information
field reading. Note, however,
that the Element Number and
Flexible
Cursor
reading cannot be output.
1003*: Actuai_Function -
outputs
in ASCII the
cursor
values
across
the entire
dis-
played trace containing the Field Selector.
1004
*: Actuai_Binary_Function - as Parameter 1003,
except
that
the
output
is in bina-
ry.
30000: the selected
text-string
is
output
as an ASCII string, followed by the selected
interface terminator. This enables,
for
instance,
programming
instructions
for
the 2032
to
be
written
into the selected
text-string
and
output
to
the 7400. A
series
of
instructions
can then be assembled on the 7400 and re-entered into
the
2032 as a
control
program.
32767: selects the
"learn"
mode
of
the 2032 allowing
output
of
a auto sequence
to
the
7400. The FIELD ENTRY knob must
not
be used when an autosequence is being
recorded. The
procedure
for
recording
an autosequence is as follows:
1.
Press SYSTEM RESET 2 and set the Interface
Terminator
to
3.
Set Special
Parameter 0
to
2032 and Special Parameter
80
to
32767.
2.
Press SETUP LABEL "Display".
* Data
output
using these modes cannot be re-entered into the 2032.