MEMORY:
(cont.)
SYSTEM RESET:
System
Reset
I
Dual Signal Enhancement
2 x
2048 sample Time Functions
2
x 2048 sample Enhanced Time Functions
Dual
Amplitude
Probability
2 x 2048 sample Time Functions
2
x 512 c.lass
Amplitude
Probability
Densities
In
addition
in all cases, the MEASUREMENT SETUP used is also
stored.
Halts a measurement
or
data
transmission
that
has been
incorrectly
set up
or
appears
to
be malfunctioning. SYSTEM RESET
stops
the
following
operations:
Recording
Averaging
Recorder
output
Digital 110
Input
Autorange
Auto
Scale
Genera~or
output
All measurement
data
and setups are retained, and an
automatic
in-
ternal DC
offset
compensation
is made.
In
addition
to
the
normal
system reset, the 2032 includes 3 higher level system reset
functions. They are selected as
follows:
1.
Press SYSTEM RESET and then a
numeric
key
2.
Release SYSTEM RESET
3.
Release the
numeric
key
The value
of
the
numeric
key can be
1,
2,
or
9. The
larger
the number, the
more
universal the system reset.
SYSTEM RESET
1:
Gives the same
action
as SYSTEM RESET, plus:
SYSTEM RESET
2:
SYSTEM RESET
9:
The signal
processor
is reset.
The
current
measured data, both instantaneous and averaged are re-
set.
The Data Source field is set to
Input.
The
echo
field is cleared.
Gives the same
action
as SYSTEM RESET
1,
plus:
The
stored
data, special parameters, and IEC SPL
program
are
reset.
The
text
strings
are cleared.
Gives the same
action
as SYSTEM RESET
2,
plus:
All Universal Parameters
are
reset.
The
Sound Intensity
Calibration
is set to:
Spacer
- 12 mm.
Temperature
-
20°C.
Pressure - 1013 mbar.
The
Sine Generator
frequency
is set
to
1000Hz.
31