Measuring Ampifier Type 2525
User Manual Vol.2
8–3
BE1394– 11
Chapter 8 – Interface Error Messages
Interface Errors
8.2 Interface Errors
Message Description Number
Header not found An unknown header was supplied to the 2525 1
Illegal data type The current interface message requires a different data type 2
Parameter error The received interface message was not unambiguously found 3
Char data not found An unknown character data was supplied to the 2525 4
Suffix data not found An unknown suffix was supplied to the 2525 5
Nr string to long The supplied numeric string is too long 6
Nr out of range The supplied numeric exceeds the legal range for the data type 7
Non-decimal out of range The supplied binary, octal or hex coded numeric exceeds the
legal range for the data type
8
String too long The supplied string is too long 9
Block data too long The supplied block data exceeds max length of block 10
Too many parameters The supplied interface message contains more parameters than
allowed
11
Out of memory System error. The 2525 was unable to allocate enough memory
to handle the interface message used. Use less complex inter-
face commands
12
Output is cancelled The response to the previous query message was not read
from the 2525
13
Deadlocked The 2525 interface is in a deadlock condition caused by input
buffer full, parser blocked, response formatter blocked and out-
put buffer full. Avoid this by always reading the response to a
query before supplying the 2525 with new interface commands
14
Unexpected byte An unexpected byte was detected when something else was
expected. For example a space was detected instead of a
message terminator
15
Interface message overrun Interface messages were received at a rate faster than the
2525 can processes them. This caused an internal input buffer
overrun
21
Interface handshake error An interface handshake error was detected. Reset the 2525
and try again
22