Measuring Ampifier Type 2525
User Manual Vol.2
Brüel & Kjær
6–28
Chapter 6 Set-up and Display (Results) Messages
Mounted Impulse/Resonance Frequency Commands
6.9 Mounted Impulse/Resonance Frequency Commands
6.9.1 Mounted_Impulse_Frequency
The amplifier uses the impulse frequency to calculate the mounted resonance fre-
quency of the accelerometer. The syntax for the command is shown in Fig.6.46.
The Frequency (Freq) is specified as follows:
Range: 3.0E+03 to 60.0E+03 Hz
Resolution: 100 Hz
Round: Nearest 100 Hz
Default: 30E+03 Hz
The syntax for the query is shown in Fig.6.47.
Fig.6.46 The syntax for the command Mounted_Impulse_Frequency
Fig.6.47 The syntax for the query Mounted_Impulse_Frequency? and amplifier response
941132e
Te
CHANNEL_1 MOUNTED_IMPULSE_FREQUENCY
To 2525
:
SP
:
Freq
NR
Hz
941133e
Te
CHANNEL_1
CHANNEL_1
MOUNTED_IMPULSE_FREQUENCY?
To 2525
:
Header OFf
Te
From 2525
: :
Value
NR
SP
:
MOUNTED_IMPULSE_FREQUENCY