BÜCHI Labortechnik AG Installation | 5
Operation Manual Vacuum Pump V-600 37/65
5.14 Operating the Vacuum Pump V-600 with 2 Rotavapor systems
It is possible to simultaneously operate 2 Rotavapor systems with only one Vacuum
Pump V-600. This involves using the alternative pump control connection (CTRL) on
the V-600, see Chapter3.2.3 "Connections", page13.
NOTE
The pump speed is then no longer controlled.
The following equipment and accessories are required:
2x Rotavapor
2x Interface I-300/I-300 Pro
2x Valve unit
2x VacuBox
1x Y-cable
2x Standard BUCHI communication cable
1x Vacuum Pump V-600
If an R-300 is not used: 2x mains adaptor and lead