EasyManua.ls Logo

Buchi V-600 - Page 46

Buchi V-600
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 | Cleaning and servicing BÜCHI Labortechnik AG
46/65 Operation Manual Vacuum Pump V-600
5 Behind the pump head is the pump diaphragm. To replace the diaphragm, see
Chapter7.2.3 "Replacing the diaphragm", page47.
Step 3: Reassembling and refitting the pump head
u Fit the pump head over the diaphragm.
u Position the fixing ring on the pump head and fix in place with four Torx-head
screws.
u Refit the connecting components between the pump heads.
u Refit the connecting component between pump head and pump intake or outlet
connection.
u Refit casing front.

Table of Contents

Related product manuals