GasAlertMax
Alarms
23
D1430/2 English
Sensor Alarm
The detector tests for a missing or defective sensor during the
activation self-test. See the section, "If the Detector Does Not
Work."
Pump Alarms
The internal pump draws air over the sensors. If the pump
stops working or the pump filter becomes clogged, the
detector activates the Pump Alarm.
Pump Alarm:
• Turn off the detector immediately
and clear the pump inlet, hose and
filter
Blocked Pump Alarm
• If the Pump Alarm continues and the
LCD advises the inlet is blocked,
see the “Pump Calibration” section.
Pump Failure Alarm: Activates if
• The pump calibration is incomplete.
• The pump is missing or broken
The pump may alarm when attaching a calibration hose,
replacing the filter, or applying calibration gas.
Low Battery Alarm
The detector tests the battery on activation and continuously
thereafter. If the battery voltage is low, the detector activates
the Low Battery Alarm.
The Low Battery Alarm continues until you replace the battery
or the battery power is almost depleted. If the battery voltage
drops too low, the detector executes an Automatic Shutdown.
Automatic Shutdown Alarm
If the battery voltage is in immediate danger of dropping
below the minimum operating voltage, the audible alarm
beeps 8 times and the visual alarm flashes 8 times. After 3
seconds, the display blanks out and the detector stops normal
operation. The display shows I LOW periodically until the
battery power is depleted.
Replace the battery. (See the section, "Replacing the
Battery.")
Note
If the Confidence Beep is on, the audible alarm does
not beep during a Low Battery alarm. (See the
section, "Confidence Beep.")
Typically, the Low Battery Alarm continues for 30
minutes before Automatic Shutdown.