2023/03 - Revision index: A - Code: 0000139233
7. Redox probe calibration
After calibration, the unit will show you the status of the probe as a %.
100%= perfect condition / 75%= good condition / 50%= fair condition, 25%= poor condition
Calibration at 1 selected point (sub-menu 1B2 "Reference" of the "Calibration" menu)
The principle is to adjust the value delivered by the instrument to that delivered by a reference redox
potential measuring instrument, when measuring the redox of the same solution.
CAL Reference
720 mV
CAL Reference
750 mV
Save and exit
5 s
Set Redox calibration
465mV
60s
H
2
O
465 mV calibration
9
10
11
4
5
6
7
1
2
3
8
The redox value measured by
the device flashes
Adjust this value to the one
read with the reference
instrument
Ex. 750 mV
Validate