12
rel. 2.7 - 16.11.2018
ENG
"Passive probes" +030220658 -
2.9 Models NTC*LT*
Storage conditions -80T105 °C
Operating range -80T105 °C
Connections M8 male-connector,
Sensor R(25 °C)= 750 Ohm 1%; Beta 3969
Precision +/- 0.2 °C at 25°C; +/- 1.15 °C from -80T105 °C
Dissipation factor (in air) 3 mW
Thermal constant over time (in air) ca. 35 s in water - 10 s in air
Cable Nickel for mounting clamp
Sensitive element index of protection IP67
Sensitive element housing Mineral oxide insulation MgO - sheath 316 - Ø 3mm radius of curvature (exception of the sensitive element)
Cable tie tension Typical 250N (position 6 on the tool); Maximum 260N (position 7 on the tool);
Classification according to protection against elec-
tric shock (sensitive element and cable)
Basic insulation for 250 Vac
Category of resistance to heat and fire Flame retardant
Insulation resistance 100 MOhm @ 500Vdc
Dielectric rigidity 1500 Vac
Maximum operating pressure: 40 bar
Tab. 2.h
Ø 4.5
6
6
6
6
M8
150 ±3 mm 26 mm
Connettore
elemento sensibile
NTC R(25°C)=750Ω
B(25/85)=3969
40 mm
Fig. 2.q
Connection cable P/Ns for NTC*LT sensor
A TSOPZCV030: silicone cable with M8 connector, 3 m long
TSOPZCV100: silicone cable with M8 connector, 10 m long
B TSOPZCV070: silicone cable extension with M8 male/female connector, 7 m long
A
Ø 9
L
Fig. 2.r
B
7000
~50
1
2
3
4
Fig. 2.s
Key:
1 M8 connector
2 rating label
3 24 AWG cable, 2 wires insulated with silicone rubber
4 M8 F connector with moulded cap