14
rel. 2.7 - 16.11.2018
ENG
"Passive probes" +030220658 -
3. NTC IMMERSION PROBE TECHNICAL SPECIFICATIONS
3.1 Models TSN1300000
Immersion probes feature the sensor directly in contact with the liquid, and are installed on the tubing. Wired using the electrical connector.
Storage conditions -40T120 °C
Operating range -40T120 °C
Sensor NTC 10 kΩ ±1% a 25 °C Beta 3435
Construction
Direct immersion with connection to the 1/8” GAS male process fitting as per UNI 338
Electrical connection
4-pin co-moulded nylon, M12x1 (DIN-VDE0627) metric thread, IP67 max. temp. 90 °C
Thermal constant over time
approx. 5 s in water - 30 s in air
Sensitive element housing AISI 316
Insulation 100 Mohm a 500 Vcc
Maximum operating pressure 40 bar
Tab. 3.a
Key:
1 NTC sensitive element 10 Kohm
2 Stainless steel socket
3 EX14
4 Co-moulded body
5 M12 male connector
55.5
1/8” GAS
13
Ø 3
1
2
3
4
5
Fig. 3.a
Accessories:
•
4-pin M12 connector for 1/8 GAS
sensor - cable length 3 m
Code TSOPZCW030
150
L=*
Fig. 3.b
Wiring:
1
2
3
NTC 10kohm
1red
2red
3 white
• 4-pin M12 connector for 1/8 GAS
sensor
code TSOPZCM000
• M12 connector can be assembled
on site, recommended cable 3x0.2
mm
2
with outer sheath.
Fig. 3.c
Wiring:
Sensor connector side view
1
2
4
3
NTC10K
• Probe socket 1/4 Gas Code
TSOPZPT000
A 1/8” GAS cyl. fitting for sensor
B 1/4” GAS cyl. process fitting
with immersion L= 10.5 mm
25.5
Ø 4
Ø 4.8
EX 17
1/4” GAS CIL.
1/8” GAS CIL.
10.5
A
B
TSN1300000
-40T120
sett/anno
Fig. 3.d
• Welding fitting Code
TSOPZCW030
13,5
17
18
3,5
Fig. 3.e