EasyManua.ls Logo

Carel NTC*HP* Series - Models PT1*HT

Carel NTC*HP* Series
76 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25
rel. 2.7 - 16.11.2018
ENG
"Passive probes" +030220658 -
Accessories
Socket: nickel-coated brass - 1413306AXX
- Maximum operating pressure 35 bar
- Temperature -20…95°C
-
60
9 11
3
Ø 8.5
CH 16
Ø 6.5
1/4 gas
Fig. 8.k
Socket 2: AISI 316 - code 1413309AXX
- Maximum operating pressure 40 bar
- Temperature -20...95 °C
60
9 11
3
Ø 7.5
CH 17
Ø 6.5
1/4 gas
Fig. 8.l
Socket: brass - 1413311AXX
56.9
15.1 11
16
R9
Ø 8.5
1/4 “ NPT thread
Fig. 8.m
Note:
cable secured with PG7 - IP68 cable gland applied to hexagonal end
kit available complete with socket and cable gland
Compression fitting with metal olive - code 1309589AXX
10
29.5
1/4” GAS Ø 6 mm AISI 316
CH 17
Ø 6
Maximum operating
pressure 40 bars
Temperature -50 to 250°C
Fig. 8.d
8.4 Models PT1*HT*
Storage conditions -50T250 °C
Operating range -50T250 °C
Connections with crimped metal terminals
Sensor Pt1000 - Class B (2 wires)
Dissipation factor (in air) ca. / approx. 7 mW/°C
Thermal constant over time (in air) ca. / approx.20 s (V=2m/s)
Cable White Teflon with two cables red and white Section. 2x0.22 mm2 Ø3 mm
Sensitive element index of protection IP67
Sensitive element housing Silicone resin
Materiale cappuccio Stainless steel Aisi 304
Dimensioni cappuccio 6x40 mm
Classification according to protection against
electric shock (sensitive element and cable)
additional insulation
Category of resistance to heat and fire Flame retardant
Resistenza Insulation 20 Mohm 500 Vcc
Dielectric rigidity 2000Vac
Tab. 8.d
40
10
5 ± 1
Ø 6
Carel code ww/yyR*
L =
(*)
(**)
Fig. 8.e
Note: (*) ww/yyR*:
ww = week of production;
yy = year of production;
R* = Revision.
(**) see table of product codes in price list