28
rel. 2.7 - 16.11.2018
ENG
"Passive probes" +030220658 -
9.2 Models TSM1500B30
Immersion probes feature the sensor directly in contact with the liquid and are secured to the tubing using a connector, available in the screw or weldable
versions.
The body is nickel-coated brass, index of protection IP67, and the gasket (O-ring) is supplied together with the probe.
Storage conditions -40T90 °C
Operating range -40T90 °C
Sensor Pt1000 Class B
Construction Direct immersion with connection to the M14 male process
Cable 2 wires AWG 22, with sheath in TPE
Thermal constant over time ca. / approx. 5 s in water - 45 s in air
Sensitive element housing Nickel-coated brass & grey PA6 co-moulded body
Insulation 100 Mohm a 100 Vcc
Maximum operating pressure 25 bar
Compatible liquids Water
Tab. 9.b
Key:
1 sensitive element
2 nickel-coated brass locking ring
3 co-moulded body
4 cable marking
5 tinned copper 2-wire cable
6 NTC sensor
7 2015 O-ring
8 nickel-coated brass thermometer
socket
23
EX14
~7.5
~40
15
Ø
3.5
Ø4
Ø 12.5
Ø 10
3
M14x1
Ø9.5
E= 2000±30
1
2
3
4
5
6
7
8
Fig. 9.f
Wiring:
NTC 10 kOhm @ 25 °C ±1%
β(25/85)
=
3435
White
Black
Fig. 9.g
PT1000Classe B

Table of Contents