26
rel. 2.7 - 16.11.2018
ENG
"Passive probes" +030220658 -
8.5 Models PT1*HF
Probe with fastening clamp
Storage conditions -50T105 °C
Sensitive element operating range -50T105 °C
Connections Stripped ends, dimensions 6±1mm
Sensor Pt1000 Class B
Precision +/- 0.8 °C; -50T90 °C
Dissipation factor (in air) 3 mW
Thermal constant over time (in air) approx. 15 s
Cable Black thermoplastic rubber flat cable (diam. 3.6x1.6 max.)
Sensitive element index of protection IP67
Sensitive element housing Thermoplastic with fastening clamp
Classification according to protection against electric shock (sensitive
element & cable)
Basic insulation for 250 Vac
Category of resistance to heat and fire UL/HB cable
Insulation resistance at 1000 Vdc >20 Mohm
Dielectric strength 1500 Vac
Tab. 8.e
20
110
6
6
L = *
cap frontal view
thermoplastic
rubber contact
cable section
Fig. 8.f
* = see table of product codes in price list
8.6 Models PT1*PS*
Storage conditions -50T105 °C
Operating range -50T105 °C
Connections Stripped and soldered ends, dimensions: 4±1 mm
Sensor Pt1000 Class B
Dissipation factor (in air) 2 mW/°C
Thermal constant over time (in air) approx. 50 min (V=1 m/s)
Cable Two-wire with double sheath, AWG22, tinned copper with resistance ≤73.9Ω/km - Insulation: TPE
specific for immersion in water on outer sheath, PP/co inside on wires, OD 3.30+/-0.10 mm
Sensitive element index of protection IP67
Case RAL7032 grey Santoprene
Food safe
Supplementary insulation for 250 Vac
Classification according to protection against electric shock
(sensitive element and cable)
Flame retardant
Category of resistance to heat and fire Flame retardant
Tab. 8.f
98
20
105 L =*
CAREL code
Magnetic base
Make a hole for fastening by screw
Fig. 8.g
* = see table of product codes in price list