Maintenance and Troubleshooting
AA2204_en | 2012-07-27 3–9
Using the Web Portal to Gain More Information on
Errors
The Web Portal is your interface to additional functions on the imager. You can view and
correct error messages and access general status information at the Web Portal.
About Codes on the Web Portal and the Display Screen
The 3-digit error and status codes on the display screen are reported at the Web Portal
as 5-digit codes. The last three digits on the Web Portal codes match the three digits on
the display screen. For example, code 701 on the display screen is the same as code 20701
on the Web Portal. Use the information in this section to understand the information at
the display screen and/or at the Web Portal and to respond appropriately.