Specifications
5–2 AA2204_en | 2012-07-27
Environmental Requirements
Temperature
• Operating: 15 to 35 °C (59 to 95 °F)
• Storage: -40 to 60 °C (-40 to 140 °F)
Relative Humidity
• Operating: 20–85 % RH, non-condensing
• Storage: 10–93 % RH, non-condensing
Altitude
30 m (100 ft) below sea level to 3,048 m (10,000 ft) above sea level
Surface Levelness
The surface where the imager is placed must be level within 1 °.
Environmental Effects
Acoustical noise:
• Less than 48 dB at 1 m during idle or standby
• Less than 75 dB momentary at 1 m during normal operation
Power Requirements
A power cord is provided with this equipment. All countries must use an
Agency-approved power cord with a plug type suitable for the country of use. Contact a
qualified dealer for help.
Connect the equipment to a power source that is suitable for the voltage and current
ratings shown on the rating label. The single-phase power source, with grounding, must
be provided within 2.5 m (8.0 ft) of the imager.
The wire must be insulation-rated for 600 V (ac). A dedicated line is recommended.
Network Requirements
The imager receives digital images from medical imaging devices (modalities) over a
10/100Base-T or 1000Base-T Ethernet Network.