CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com 22
5.9 Limitation to ZERO and TARE operation under different conditions:
Table5-1
Standard
Weight
on
platform
Data in TARE
memory unit
key function
Tare key Zero key
USA
0
no No action
Zero
yes Clear the tared weight
0
No
Tare
Yes
Canada
0
no No action
Zero
Yes Clear the tared weight
0
No Tare
Yes No action
Europe
0
No No action Zero
Yes Clear the tared weight
zero and clear the
tared weight
0
No
Tare
zero
Yes
zero and clear the
tared weight
None
(same with
Europe)
0
No No action Zero
yes Clear the tared weight
zero and clear the
tared weight
0
No
Tare
zero
yes
zero and clear the
tared weight
NOTE: (1) ZERO only be active when scale is stable and weight is in SAZSM setting range.
(2) TARE only be active when scale is stable
(3) Clear TARE weight or ZERO scale, make indicator to enter displaying GROSS mode
(4) TARE weight, make indicator to enter displaying NET mode
5.10 Output data: When scale is stable, press PRINT key.
5.11 Accumulation:
Press ACC key to add displayed number to accumulation memories, and accumulation times will also
add up 1. and then to display accumulation result
5.12 Change Weight UNIT:
Short press UNIT key to select kg, lb, lb:oz, g, or oz unit, note: under some condition, g and lb:oz is not
available. And , In trade application, lb:oz should be prohibited. Please refer the following tables: