CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com 28
9. HOLD Function
NOTE: In trade application, HOLD function may not be allowed, please check with your local
legal organization of measurement!
9.1 HOLD function can be used to freeze display number. In this mode, scale can catch a dynamic
number, hold a stable number, or average a unstable number, then HOLD (freeze) this number
temporary for user to watch or record. This function can be used in normal weighing mode, counting
mode and percent weighing mode. After entering HOLD mode, the speed of A/D converter can be
increased to 80Hz (if CONFIG-AD.H.SPD is set to YES) from original 10Hz for some dynamic
weighing applications. With the hold function, it is possible to weigh restless weighing samples such
as live animals, moving objects. The indicator provides special mode settings to accommodate
sample’s movements.
9.2 To make HOLD function be active, the CONFIG-FUNC-HOLD menu item must be set to YES; menu
items of USER-HOLD-HLD.MOD /-AVG.TIM /-HLD.TIM /-HLD.RNG /-STB.TIM,
USER-OTHER-NLD.RNG need be set to reasonable value.
To speedup sampling of weight, set CONFIG-AD.H.SPD menu item to YES.
To enter HOLD working mode, press down HOLD key when scale works in normal weighing mode,
counting mode or percent weighing mode.
9.3 There’re several HOLD mode to freeze display data:
(1) Positive Peak Number HOLD mode
(2) Negative Peak Number HOLD mode
(3) Toggle HOLD mode
(4) Average HOLD mode
(5) Auto HOLD mode
The following are details of these HOLD modes:
9.3.1 Positive Peak HOLD:
When USER-HOLD-HLD.MOD is set to PS.PEAK, the hold mode is positive peak hold mode. When
scale first enters this working mode, it will display the largest positive number that is from the time of
zero-pint set. After entering this working mode, scale will always catches and refresh positive larger
number and display it. To exit HOLD mode, press HOLD key again.
9.3.2 Negative Peak HOLD:
When USER-HOLD-HLD.MOD is set to NG.PEAK, the hold mode is negative peak hold mode. When
scale first enters this working mode, it will display the largest negative number that is from the time of
zero-point set. After entering this working mode, scale will always catches negative larger number and
display it. To exit HOLD mode, press HOLD key again.
9.3.3 Toggle HOLD:
When USER-HOLD-HLD.MOD is set to TOGGLE, the hold mode is toggle hold mode ---a manual Hold
function. After entering this working mode, scale will freeze and display number if scale is stable. Only

Related product manuals