CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com 29
the weight that is over USER-OTHER-NLD.RNG (zero 'dead' band ) can be held. To exit HOLD mode,
press HOLD key again. If the time of scale being unstable is more than USER-HOLD-STB.TIM, STB.ER
will be shown, press TARE key to start averaging again, or press HOLD key to exit.
9.3.4 Average HOLD:
When USER-HOLD-HLD.MOD is set to AVERAG, the hold mode is average hold mode. After entering
this working mode, scale will freeze and display number if scale is stable. If scale is not stable, but the
variation is less than USER-HOLD-HLD.RNG, scale will average data in USER-HOLD-AVG.TIM, then
freeze and display the number. Only the weight that is over USER-OTHER-NLD.RNG can be frozen.
Scale will exit HOLD mode according to the setting of USER-HOLD-HLD.TIM. If the time of scale
variation being over USER-OTHER-NLD.RNG is more than USER-HOLD-STB.TIM, STB.ER will be
shown, press TARE to start averaging again, or press HOLD key to exit.
9.3.5 Auto HOLD:
When USER-HOLD-HLD.MOD is set to AUTO, the hold mode is auto hold mode--- different subjects
can be weighed one after another without pressing any buttons. After entering this working mode, scale
will freeze and display number if scale is stable. If scale is not stable, but the variation is less than
USER-HOLD-HLD.RNG, scale will average data in USER-HOLD-AVG.TIM, then freeze and display the
number. Only the weight that is over USER-OTHER-NLD.RNG can be frozen. If the held weight is
moved away, and a new load put on the scale, scale will automatically hold new number of load. Scale
will exit HOLD mode according to the setting of USER-HOLD-HLD.TIM. If the time of scale variation
being over USER-OTHER-NLD.RNG is more than USER-HOLD-STB.TIM, STB.ER will be shown,
press TARE to start averaging again, or press HOLD key to exit.
9.4 In Positive or Negative Peak HOLD mode, the red HOLD (for LED version) or PEAK and HOLD (for LCD
version) announciator will be lighted, in other HOLD mode, green HOLD (for LED version) or HOLD (for
LCD version) announciator will be lighted. When HOLD announciator flash, the displayed number is live,
When HOLD announciator become steady, the displayed number is frozen.
10. Data Compare Function
10.1 Data compare function can be used in normal weighing mode, counting mode and percent weighing
mode, and call it as Check Weight, Check Counts and Check Percentage. When this function is enabled,
you can set a higher and a lower limitation of weight, counts or percentage independently, and these
limitation can be saved permanently. Then, the current data of weight , counts or percentage will be
compared with the setting limitation, and corresponding annunciator will be lighted.
10.2 To make data compare function be available, CONFIG-FUNC-COMPAR menu item should set to YES,
and high and low limitation should be set correctly according to following steps:
10.3 In normal weighing mode, counting mode or percent weighing mode, Press down DATA key more than
3s to enter input compare data of high and low mode.
10.4 After HIGH being shown, last setting data of high will be displayed, use PRINT, ACC, UNIT key to input
new number of high and press TARE key to confirm. Annunciator of Hi will be shown in this step. Press
ZERO key to exit and back to original working mode.