23.9 Communication Details when USER-COM1/2-LAYOUT is set to SCP-12:
23.9.1 This protocol of serial communication is similar to with NCI3835 protocol. The baud rate and data
format is set by User menu.
23.9.2 Output status bit meaning:
Bit Status Byte1 Status Byte2
0
0=Scale in motion 1 = Under capacity
1=Stable 0 = Not under capacity
1
0= Scale at zero 1 = Over capacity
1= Not at zero 0 = Not over capacity
2
0=RAM error 1 = ROM error
1= RAM okay 0 = ROM okay
3
0= EEPROM error 1 = Faulty calibration
1= EEPROM okay 0 = Calibration okay
4
Always 1 Always 1
5 always 1 always 1
6 always 0 always 0
7 parity parity
23.9.3 Key to symbols used:
<ETX> End of TeXt character (03 hexadecimal).
<LF> Line Feed character (0A hex).
<CR> Carriage Return character (0D hex).
Xxxxxx Weight characters from display including minus sign and out-of-range characters.
p Polarity character (ie ‘-’ for negative, space for positive)
hh Two status bytes. (see 23.9.2)
UU Units of measure (LB, KG or OZ all upper case).
23.9.4 Summary of Command and Response:
Command
Response
ASCII HEX
W<CR> 57 0D
Returns decimal lb, kg or oz weight, units and status.
<LF>pxxx.xxUU<CR>hh<ETX>
Returns ounces weight with units plus scale status.
<LF>p00xxxxxOZ<CR>hh<ETX>
Scale status only if initial zero error.
<LF>hh<CR><ETX>
S<CR> 53 0D Read scale status : <LF>hh<CR><ETX>
Z<CR> 5A 0D Simulate ZERO key: no response from scale.
others
Un-known commands: <LF>?<CR>
23.9.5 If your indicator need work with UPS worldship, you can try following settings:
(1) USER-COM1(or 2)-BAUD.RT=4800
(2) USER-COM1(or 2)-BYT.FMT=7E1
(3) USER-COM1(or 2)-LAYOUT=SCP-12
(4) Set scale port to NCI3835 in UPS Worldship.