EasyManua.ls Logo

CAS CI-100A - 21. Serial Port1;2 (COM1;2) Receiving Test; COM1;2 Receiving Test Setup and Execution

CAS CI-100A
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com 43
23.9 Communication Details when USER-COM1/2-LAYOUT is set to SCP-12:
23.9.1 This protocol of serial communication is similar to with NCI3835 protocol. The baud rate and data
format is set by User menu.
23.9.2 Output status bit meaning:
Bit Status Byte1 Status Byte2
0
0=Scale in motion 1 = Under capacity
1=Stable 0 = Not under capacity
1
0= Scale at zero 1 = Over capacity
1= Not at zero 0 = Not over capacity
2
0=RAM error 1 = ROM error
1= RAM okay 0 = ROM okay
3
0= EEPROM error 1 = Faulty calibration
1= EEPROM okay 0 = Calibration okay
4
Always 1 Always 1
5 always 1 always 1
6 always 0 always 0
7 parity parity
23.9.3 Key to symbols used:
<ETX> End of TeXt character (03 hexadecimal).
<LF> Line Feed character (0A hex).
<CR> Carriage Return character (0D hex).
Xxxxxx Weight characters from display including minus sign and out-of-range characters.
p Polarity character (ie ‘-’ for negative, space for positive)
hh Two status bytes. (see 23.9.2)
UU Units of measure (LB, KG or OZ all upper case).
23.9.4 Summary of Command and Response:
Command
Response
ASCII HEX
W<CR> 57 0D
Returns decimal lb, kg or oz weight, units and status.
<LF>pxxx.xxUU<CR>hh<ETX>
Returns ounces weight with units plus scale status.
<LF>p00xxxxxOZ<CR>hh<ETX>
Scale status only if initial zero error.
<LF>hh<CR><ETX>
S<CR> 53 0D Read scale status : <LF>hh<CR><ETX>
Z<CR> 5A 0D Simulate ZERO key: no response from scale.
others
Un-known commands: <LF>?<CR>
23.9.5 If your indicator need work with UPS worldship, you can try following settings:
(1) USER-COM1(or 2)-BAUD.RT=4800
(2) USER-COM1(or 2)-BYT.FMT=7E1
(3) USER-COM1(or 2)-LAYOUT=SCP-12
(4) Set scale port to NCI3835 in UPS Worldship.

Related product manuals