EasyManua.ls Logo

CAS CI-100A - 23. Serial Communication Details; COM Ports, Settings, and String Format

CAS CI-100A
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com 45
24.5 SIO---Serial Input Output Connector
PIN # DEFINITION IN/OUT/POWER ELECTRICAL LEVEL
1 RS485 signal A (if RS485 installed) Input/output 0-5Vdc
2 RS232 Transmit on COM1 Output -12 to +12Vdc
3 RS232 Receive on COM1 Input -12 to +12Vdc
4
5 GND Power ground 0Vdc
6
7
8
9 RS485 signal B (if RS485 installed) Input/output 0-5Vdc
24.6 J1---USB Connector for virtual RS232 #1 and USB power supply
PIN # DEFINITION IN/OUT/POWER ELECTRICAL LEVEL
1 VDD Power output
5±0.3 Vdc
2 RXD Receive on COM2 Input 0-5Vdc
3 TXD Transmit on COM2 Output 0-5Vdc
4 GND of VDD Power ground 0Vdc
5 GND1 of VUSBH Power ground 0Vdc
6 USB Power DC/DC select Output 0-5Vdc
7 USB Power DC/DC output output
6±0.3 Vdc
24.7 KCAK Jumper set:
CONNECTED PINS FUNCTION
1-2 Calibration Disabled
2-3 Calibration Enabled
24.8 JP1 Jumper:
CONNECTED PINS FUNCTION
1-2 Two shorter on pin1-2: 4 wires Load cell is used
2-3 Two shorter on pin2-3: 6 wires load cell is used
24.8 ICM1 Socket: used for optional Bluetooth, RS485 or USB interface
24.9 JTAG and IOSEL used by Manufacture